Individual channel modules that plug into the 3 or 16-channel recorder condition, amplify, filter, digitize and store the transducer signals. Simultaneous sampling provides excellent time correlation between channels and external events. The digitizer is available with 14, 16 or 24-bit resolution and sample rates up to 10 Million samples per second.

Each channel has two inputs: one for bridges, voltage, RTDs, potentiometers and thermocouples; the other AC coupled with current excitation for transducers with built-in electronics. Signal conditioning includes programmable excitation and completion for ¼, ½ and full bridge, AC-coupling, automatic balance, voltage and four-step shunt calibration, programmable gain and a four frequency, eightpole filter. High-level analog outputs are available.

A high-level instruction set is provided for user programming or PI580, a Windows application, may be used. Operating parameters are stored in non-volatile memory and automatically loaded when the system is powered up or reset. Learn more about PI580 Software.

Features Model 5841 Model 5842 NEW Model 5843
Transducer Excitation 0 to 12 Volts,
50 mA.
0 to 12 Volts,
50 mA.
0 to 12 Volts,
50 mA.
Bridge Input 4-Wire, programmable for 120 or 350 Ohm 1/4, and all 1/2 and full bridges.
Automatic balance.
4-Wire, programmable for 120 or 350 Ohm
1/4, and all 1/2 and full bridges.
Automatic balance.
4-Wire, programmable for 120 or 350 Ohm 1/4, and all 1/2 and full bridges.
Automatic balance.
Preamp Input AC coupled, 1 to 20 mA power. AC coupled, 1 to 20 mA power. AC coupled, 1 to 20 mA power.
Calibration Voltage substitution,
4-step shunt calibration.
Voltage substitution,
4-step shunt
calibration.
Voltage substitution,
4-step shunt calibration.
Amplifier Differential input, programmable gain 1 to 1,000. Differential input, programmable gain 1 to 1,000. Differential input, programmable gain 1 to 1,000.
Bandwidth 100 kHz 300 kHz 100 kHz
Filter Four-frequency, eight-pole Bessel low-pass, 5 kHz,
10 kHz, 20 kHz,
50 kHz.
Four-frequency, eight-pole Bessel low-pass, 10 kHz,
20 kHz, 50 kHz,
100 kHz.
Four-frequency, eight-pole Bessel low-pass, 5 kHz,
10 kHz, 20 kHz,
50 kHz.
ADC 16-bits 14-bits 24-bits
Sample Rate 10, 20, 50, 100S/s
......1 MS/s
20, 50, 100, 200S/s
......2 MS/s
1K, 2K, 5K, 10K, 20K , 50KS/s ......200KS/s
Data Storage (Words/Ch.) 1M, 2M or 4M with 40 hour backup 1M, 2M or 4M with 40 hour backup 1M, 2M or 4M with 40 hour backup