  Model 6132 - 4-Ch. Transducer Amplifier

FEATURES

  Programmable excitation, remote sensing
  Programmable input configuration
  Shunt and voltage calibration
  Automatic zero and balance
  Gains 1 to 5,000 with 0.05% accuracy
  Continuous 10 Volt analog outputs

DESCRIPTION

The 6132 input module has four channels of high performance signal-conditioning amplifiers for strain gages and bridge transducers. Each channel has programmable excitation with remote sensing, voltage calibration, local or remote shunt calibration, programmable gain instrumentation amplifier and four-pole low pass filter.

The 6132 is used with quarter, half and full bridge transducers, potentiometers and low-level voltage signals in demanding applications such as load control. The EM option adds continuous excitation monitoring with out-of-limit alarms. The PF option adds a four-pole, 4 to 1,000 Hz programmable filter with 1 Hz resolution.

Voltage substitution using an external voltage standard is provided for traceable gain calibration. Internal or external shunt calibration is provided for transducer calibration. Transducer balance, zero and gain calibration are automatic. Two programmable alarms with upper and lower limits are checked for each output. The highlevel analog outputs provide a means to independently monitor or record each channel.

Specifications - Model 6132 - 4-Ch. Transducer Amplifier
Unless noted specifications apply to each channel individually
Input Two to eight wires with shield. Bridge configuration is programmable for ¼, ½ and full bridge, 120 Ohm and 350 Ohm.
Bridge Excitation Programmable from 0 to 12 Volts with 3.3 mV resolution and ±0.01% regulation, 50 mA limited to 70 mA.
Current Excitation Programmable from 0 to 50 mA with 12 uA resolution and ±0.01% regulation, 10 Volt compliance.
Balance Automatic to ±0.05% of range ±1 mV. Range up to 10 mV/V, 2 mV/V (350 Ohms) installed. Stability is ±0.02%, ±0.005%/°C.
Shunt Calibration Single step shunt, either polarity, internal or external connection, 0.502 mV/V (350 Ohm bridge), ±1% installed.
Voltage Calibration Alternate input for external calibration source. Programmable attenuator with steps of 1, 0.1 and 0.01, ±0.02% accuracy. Output of the attenuator is provided for calibration.
More Specs Read more about Specifications in the 6132 Data Sheet.