  Model 6060 - 2-Channel Transducer Amplifier-Digitizer 300 Volts Common Mode

FEATURES

  Voltage & current excitation with remote sensing
  Isolated excitation and input
  300 Volts common mode
  Gains 1 to 5,000
  50 kHz Bandwidth, 100 kHz optional
  Voltage & shunt calibration
  Analog outputs

DESCRIPTION

The 6060 is a two-channel signal conditioning amplifier-digitizer module with 50 kHz or 100 kHz bandwidth and both digitized and analog outputs. The bridge input is ten-wire shielded with programmable constant voltage or constant current excitation and programmable completion for quarter, half and full bridge transducers. Automatic bridge balancing accommodates large unbalances without limiting dynamic range.

The input and excitation are isolated from the outputs, power and control interface. This gives the user complete freedom to ground the input without creating ground loops that introduce noise and offset errors.

The differential instrumentation amplifier has programmable gains from 1 to 5,000 and automatic zero. The standard filter is a six-pole Bessel with four programmable bandwidths. An optional four-pole Bessel filter has continuously programmable bandwidth from 4Hz to 20Hz.

A “features card” provides shunt calibration using dedicated inputs. Two-step, resistive shunt calibration is standard. Four step shunt calibration and DAC shunt with 16-bit resolution are also available. Voltage substitution using an external reference is provided for traceable gain calibration.