Model 6014

FEATURES

  Excitation for current driven transducers
  Gains 1 to 5,000 with 0.05% accuracy
  AC or DC input coupling
  Automatic zero
  Four-pole, low pass filter
  Simultaneous sample & hold
  16-bit ADC resolution
  Programmable warnings and alarms

DESCRIPTION

The 6014 has eight channels of AC or DC coupled programmable gain instrumentation amplifier, filter and sample and hold. A high-level multiplexer selects each channel for digitizing and output to the 6000 data bus. It includes constant current excitation for use with current driven transducers. The 6014 is primarily for use with transducers that have a built-in, low-impedance output amplifier or charge to voltage converter. It may also be used with voltage inputs where AC coupling is desired . AC or DC coupling and current excitation are selected by jumpers for each channel. A continuous analog output is available for monitoring and output to tape recorder or display.

Specifications - Model 6014 8-Channel Accelerometer / Transducer Digitizer
Unless noted, specifications apply to each channel individually
Excitation 6 mA, ±15%, up to 28 VDC compliance, jumper enabled (requires enclosure power option).
Input Two wire with shield; jumper selected AC or DC coupling.
Calibration Voltage substitution with attenuation of 1, 0.1 and 0.01.
Range ±2 mV to ±10 Volts. Limited bandwidth from ±2 mV to ±50 mV. Input protected to ±50 Volts
Gain Programmable from 1 to 5,000 in 1, 2, 3, 5, 10 steps with ±0.05% accuracy.
Static Accuracy Typically ±0.02% for gains 1 to 1,000 following calibration.
Common Mode 110 dB @ gain 1,000, ±10 Volts, protected to ± 50 Volts.
Bandwidth 5 kHz for gains 1 to 100. 1 kHz for gains >100. Four-pole low-pass filter with plug-in modules for 10 Hz to 5 kHz, 1 kHz supplied. AC coupled high-pass is <3 Hz.
Sampling Up to 10k samples per second per channel. Sample & hold time correlation is ±50 nS.
ADC 16-bits, two's complement output.
Analog Output ±3 Volts full scale.