Pacific Instruments
Call: 925-827-9010
Contact Us Login: Client Home
  Model 6005 DC Operated, Ruggedized Portable Mainframe
DC Operated, Portable Data Acquisition Mainframe
Model 6005

FEATURES

  Portable, DC Operated, AC Power Adapter Available
  Ruggedized for Mobile Applications
  4 Million Samples Per Second using USB Interface
  IEEE-488 Interface with Transient Capture
  PCM Telemetry and SCRAMNet Data Output Options
  Fast, Hardware-Based Alarms with Digital Outputs
  All Front-Mounted Controls and Connectors

DESCRIPTION

The 6005 Mainframe is DC operated and ruggedized for mobile systems. The USB version with ten input/output slots accommodates up to 80 strain gage or bridge transducers, thermocouples, LVDTs, RTDs, voltage or current channels with amplifier per channel architecture. Systems can also be configured with test sequencers, digital inputs and outputs and a DSP control processor that can implement PID control loops and display and record derived measurement parameters.

This mainframe supports all Series 6000 signal conditioning and I/O modules. All connections are made on the front allowing it to be located in the tightest of spaces. It has integral fans that supply cooling air to the modules and power supply.

The 10-slot 6005U has a USB 2.0 control and data interface that achieves data rates in excess of 4 million samples per second and is compatible with the USB ports on most computers including laptops. Like other 6000 mainframes, it supports systems of up to 4096 channels using slave enclosures.

The 9-slot 6005 mainframe with IEEE-488 GPIB interface achieves data rates to 800K samples per second. It includes a 2 million-word ring buffer for event data capture and supports PCM and SCRAMNet data output options. PCM output is IRIG 106, Ch. 4 compatible and SCRAMNet provides high-speed, low-latency, cross-platform data output.

Each input/output module requires a connector panel when used in the 6005 Mainframe enclosure. These panels provide input and output signal connections on the front of the enclosure to facilitate wiring and cable routing in compact installations.

PDF  For complete Specifications, see 6005 Data Sheet (PDF)

Specifications
Model 6005
Model 6005U
Model 6005UE
Capacity Up to 9 Series 6000 input and output modules. Up to 10 Series 6000 input and output modules. Up to 10 Series 6000 input and output modules.
Data 16-bits, 2's complement, formatted in two bytes, low-byte followed by high-byte. 16-bits, 2's complement, formatted in two bytes, low-byte followed by high-byte. 16-bits, 2's complement, formatted in two bytes, low-byte followed by high-byte.
Sample Rate Programmable with 1 µS resolution up to 1 Million samples per second. Programmable with 1 µS resolution up to 4 Million samples per second. Programmable with 1 µS resolution up to 4 Million samples per second.
Interface IEEE-488.2 with high speed handshake. USB 2.0. USB 2.0 or Ethernet with remote control capability.
Power 10 to 20 VDC. (21 to 32 VDC and 21 to 56 VDC also available) 10 to 20 VDC. (21 to 32 VDC and 21 to 56 VDC also available) 10 to 20 VDC. (21 to 32 VDC and 21 to 56 VDC also available)
Environmental 0°C to +50°C operating. Humidity to 95% without condensation. 0°C to +50°C operating. Humidity to 95% without condensation. 0°C to +50°C operating. Humidity to 95% without condensation.
Size/Weight 13.4 inches wide, 10.5 inches high, 16.7 inches deep. Approximately 30 pounds with all channel modules. 13.4 inches wide, 10.5 inches high, 16.7 inches deep. Approximately 30 pounds with all channel modules. 13.4 inches wide, 10.5 inches high, 16.7 inches deep. Approximately 30 pounds with all channel modules.
Model 6005 Applications
6005 Portable DAQ System in a Jeep 6005 with a Laptop &
PI660 Application Software
is a Complete, Portable Data Acquisition & Recording System
  • Structural Testing
  • Automotive Testing
  • Earthquake Engineering
  • Wind Tunnels
  • Engine Testing
  • Mine Safety
  • Locomotive & Rail Car
  • Rocket Motor
  • Launch Monitoring
  • Explosive Effects
  • Vibration Monitoring
  • Nuclear Propulsion
  • Ship Hull & Equipment
  • Power Generation
  • Power Distribution
  • Truck & Bus Chassis
  • Flight Testing
  • Military Vehicles
  • Stress Analysis
  • Materials Testing
  • Ordering Information
    6005 Mainframe, 12 VDC, 9-slot, IEEE-488 data and control interface.
    6005U Mainframe, 12 VDC, 10-slot, USB 2.0 data and control interface.
    6005UE Mainframe, 12 VDC, 10-slot, USB 2.0 or Ethernet data and control interface.
    6005P Mainframe, 9-slot, IEEE-488 data and control interface, IRIG 106, PCM data output.
    6006 Slave enclosure, 9-slot.
    6085 Connector Panel, specify input/output module type. Example: 6085-6013.
    ©2009 Pacific Instruments | Home | Products | Support | News | About Us | Sitemap